18
The following table is a typical temperature coefficients (percentage of change of
conductivity per).
Solution
% /
C
℃
Ultra pure Water
4.55
Salt(NaCl)
2.12
5% NaOH
1.72
Dilute Ammonia
1.88
10% HCl
1.32
5% Sulfuric Acid
0.96
98% Sulfuric Acid
2.84
Cell Constant
For conductivity measurement of a solution, can accurately measure by adjusting cell
constant.
Cell constants consist of 0.01, 0.1, 1.0, 10 and 100, and set
by using
or
key.
Standard Solution
After setting cell constant, press
Setup
key, and then the display is shown as follows.
If pressing
Se`lect
key, conductivity of standard solution
(146.9
S, 1413
S, 6.67mS, 12.9mS, 111.9mS) is displayed
in turn.
If using standard solution that not showing on screen, adjust
conductivity by using
or
key.
Chapter IV Setup Functions
N eo M et
SETUP
1.0
SETUP
1413
S