26
2)
Calibration of CAL2 to CAL5 (Buffer 2 to Buffer 5)
Clearly rinse electrode and put into the second buffer. The
calibration method of CAL2~CAL5 is the same as done in
CAL1. If finishing calibration, is changed into the initial
display automatically.
If pressing
Cal
key after calibrating 2 points, change to the
initial display.
Put electrode into sample, and press
M easure
key.
The measured data is stored by pressing
M em ory
key manually. Refer to Data-Log of
Chapter VI.
Chapter V Calibration and Measurement
N eo M et
CAL 2
1.00
ATC
25.0
C
ISE
10
2
5.07
ATC
25.0
C
ISE
10
1