16
ISE M ode Setup
Standard Solution Setup
In pH ready condition press Mode key to enter ISE mode. In
ISE ready condition if pressing Setup key, the display is
shown as follows.
If pressing
Select
key, concentration of buffer (1.00
10
-2
,
1.00
10
0
, 1.00
10
1
, 1.00
10
2
, 1.00
10
3
) is played in turn
and then set buffer by using
M emory
key. It is available to
set buffer up to 5 points. Please note that it is not available to
set just only 1 point.
If finishing setup, press
O ut
key exit setup mode.
DO/O
2
/Air/TEMP Meter
DO M ode Setup
Altitude
In the initial display of DO, press
Setup
key to enter Setup and then the display is
shown as follows.
Adjust altitude with
or
key.
The set altitude is automatically compensated.
If finishing the setting of conditions, press
O ut
key to exit or
Setup
key to enter the next setup.
Salinity
Adjust salinity with
or
key.
The set salinity is automatically compensated.
If finishing the setting of conditions, press
O ut
key to exit or
Setup
key to enter the next setup.
Chapter IV Setup Functions
N eo M et
SETUP
CAL 1
1.00
ISE
10
2
SETUP
0
ALT
SETUP
0.0
Sal