Invacare® Platinum™ Mobile
Conserver Trigger
Sensitivity:
< 0.18 cmH
2
O pressure drop (For all cannula lengths)
Factory set—no adjustment, pressure activated
Only patient respiratory efforts that achieve the trigger pressure will result in the delivery of an oxygen bolus.
Conserver Breath Rate
Capacity:
15–40 BPM (breaths per minute) without reduction of bolus minute volume
Maximum Outlet Pressure:
28.5 psig (197 kPa)
Cannula Requirements:
Length: 4–25 ft (1.2–7.6 m) including all oxygen tubing
Tubing: crush-proof, single lumen
Adult, standard flow (rated for up to 6 L/min continuous flow) for lengths up to 7 ft)
Adult, high flow (rated for up to 15 L/min continuous flow) for lengths greater than 7 ft to 25 ft)
Example of Possible Cannula Model: Westmed Inc. Part Number 0194 (4 ft length)
Battery Specifications
(each battery):
Rechargeable lithium-ion, 14.4 V, 5800 mAh, 83.5 Wh, 500 full charge/discharge cycle life
Battery Shelf Life:
12 months from date of manufacture
Condition
One Battery
Two Batteries
Pulse Setting P1
5 hr 5 min
10 hr 10 min
Pulse Setting P2
3 hr 30 min
7 hr 0 min
Pulse Setting P3
2 hr 20 min
4 hr 40 min
Battery Duration:
(Times are approximate)
Pulse Setting P4
1 hr 45 min
3 hr 30 min