Invacare® Platinum™ Mobile
Concentrator is Beeping with No Lights or Display
Symptom
Probable Cause
Solution(s)
Battery has been removed or disconnected and
there is no external power connected.
1.
Ensure the battery is completely inserted and
restart.
2.
Install a different charged battery and restart.
Concentrator is beeping with no
lights or display.
Beeping will last less than
two minutes if no power
source is provided.
When operating without a battery, external power
is lost.
1.
Try another power outlet and check connections
to concentrator and restart.
Battery Not Charging When External Power is Connected
Symptom
Probable Cause
Solution
Power source is faulty, or there is a loose connection.
1.
Try another power outlet and check connections to
concentrator.
Battery is outside the allowed temperature range for
charging.
1.
Allow concentrator and battery to cool down to less
than 104°F (40°C), or warm up to 41°F (5°C).
Battery is not fully seated.
1.
Ensure the battery is completely inserted.
With AC or DC
power plugged into
the concentrator,
the external power
icon is illuminated
and battery charge
level indicator is not
flashing.
Battery charge is complete (all four segments of charge
level indicator are illuminated).
1.
No action required.