Troubleshooting
LCD DISPLAY TEXT:
DESCRIPTION:
SOLUTIONS:
Shut Down
System Fault
INDICATORS:
Two audible beeps every 16 seconds
YELLOW alarm indicator illuminated
Abnormal system condition has been detected.
May be caused by malfunction of compressor,
fan, pressure or software.
Concentrator function has shut down.
1.
Turn the concentrator Off, then On again
to retry.
2.
Change to another source of oxygen if
alarm continues
NOTES:
To clear the alarm, turn the concentrator off. Restart concentrator.
Call your provider and report the problem if alarm continues.
LCD DISPLAY TEXT:
DESCRIPTION:
SOLUTIONS:
Not applicable. All power lost, display is blank.
INDICATORS:
Audible beeps every second for approximately
30 to 60 seconds
No other control panel indicators illuminated
All power sources have been lost/removed.
Concentrator function has shut down.
1.
Ensure the battery is completely inserted.
2.
Check external power connections to the
concentrator and to the power outlet.
3.
Try another power outlet.
4.
Change to another source of oxygen if
alarm continues.
NOTES:
This alarm occurs even if the concentrator is off when all power is removed.
To clear the alarm, complete solutions 1, 2, or 3 before restarting.
1194969-A-00
59
Содержание Platinum Mobil
Страница 71: ...Notes ...