Troubleshooting
7.2.2
Shut Down Alarms
The concentrator shuts down when the alarm conditions in this section occur.
LCD DISPLAY TEXT:
DESCRIPTION:
SOLUTIONS:
Shut Down
No Breath Detected
Oxygen concentrator has not detected a
breath for over a two minute period.
1.
Verify cannula is connected, not kinked,
properly positioned and user is breathing
through his/her nose.
2.
Change to another source of oxygen if
alarm continues.
NOTES:
This alarm does not apply during warm up.
To clear alarm, turn the concentrator off. Complete solution 1 before restarting. Press and hold the On/Off button to power down
the concentrator and restart.
LCD DISPLAY TEXT:
DESCRIPTION:
SOLUTIONS:
Shut Down
Stuck button
INDICATORS:
Two audible beeps every 16 seconds
YELLOW alarm indicator illuminated
A stuck button is detected upon connection
to power source.
1.
Remove all power sources from
concentrator. Press each control button
looking for a stuck button. Reconnect
power.
2.
Change to another source of oxygen if
alarm continues.
NOTES:
To clear alarm, turn the concentrator off. Complete solution 1 before restarting. Call you provider and report the problem if alarm continues.
1194969-A-00
55