Troubleshooting
No External Power Symbol Illuminated on Control Panel
Symptom
Probable Cause
Solution
With AC power
plugged into
concentrator,
external power icon
is not illuminated on
the control panel.
Power source is faulty, or there is a loose connection.
1.
If the green indicator light on the AC adapter is lit,
check power connections from the AC adapter to
concentrator.
2.
If the green indicator light on the AC adapter is not lit,
check power connections from outlet to AC adapter
or try another outlet.
Power source is faulty or there is a loose connection.
1.
Check power connections to concentrator.
2.
Try another power outlet.
With DC power
plugged into
concentrator,
external power icon
is not illuminated on
the control panel.
The DC power cable fuse has blown.
1.
Switch to AC power and contact your provider to
service the DC power cable.
7.2
Alarm Conditions
When any alarm condition occurs, the yellow alarm indicator light on the control panel illuminates and text describing the alarm condition
is shown on the display screen. Refer to the solution section of the alarm tables for potential corrective actions. If necessary, contact
your provider.
In all cases, pressing and holding the On/Off button for one second will turn off and reset the concentrator.
If the cause of the alarm condition is not corrected, it will reappear when the concentrator is turned back on.
All alarms are classified as Low Priority technical alarm conditions.
When multiple alarm conditions exist, the text associated with the highest ranked alarm will be shown on the display screen.
7.2.1
Operational Alarms
The concentrator continues to run when the alarm conditions listed in this section occur.