Troubleshooting
LCD DISPLAY TEXT:
DESCRIPTION:
SOLUTIONS:
Cannot charge
Allow to warm
INDICATORS:
Single audible beep every 30 seconds
YELLOW alarm indicator illuminated
Battery is too cold to allow the battery to
charge.
This alarm only occurs when an external
power source is connected.
1.
Move concentrator to warmer
surroundings. Allow concentrator to
warm up to 41°F (5°C), OR
2.
Remove battery (with AC or DC power
connected) to allow battery to warm.
NOTES:
The alarm will turn off when the required battery temperature is reached, when the battery is removed, or when external power is disconnected.
LCD DISPLAY TEXT:
DESCRIPTION:
SOLUTIONS:
Sensor Fault
See Manual
INDICATORS:
Single audible beep every 30 seconds
YELLOW alarm indicator illuminated
Oxygen sensor feedback is out-of-range.
Sensor failure is likely.
1.
Turn concentrator Off, then On again to
retry.
NOTES:
The alarm will turn off when the sensor feedback is back in range.
Call your provider and report the problem if alarm continues.
1194969-A-00
53
Содержание Platinum Mobil
Страница 71: ...Notes ...