SYMPTOM
PROBABLE CAUSE
SOLUTIONS
Battery indicator flashes the charge level
is low - immediately after recharge.
Battery failure.
Malfunctioning
battery charger.
Electrical
malfunction.
Check batteries for shorted cell. Replace if necessary.
Contact Provider/Invacare for Service.
Contact Provider/Invacare for Service.
Battery indicator flashes the charge level
is low - too soon after being recharged.
Batteries not
charged.
Weak batteries.
Have charger checked.
Replace batteries if necessary. Refer to
Motor “chatters” or runs irregular.
Motor/gearbox
malfunction.
Stop use of Wheelchair. Contact Provider/Invacare for Service.
Joystick erratic or does not respond as
desired.
Damaged motor
coupling.
Electrical
malfunction.
Controller
programmed
improperly.
Contact Provider/Invacare for Service.
Contact Provider/Invacare for Service.
Contact Provider/Invacare to have controller reprogrammed.
Wheelchair does not respond to
commands.
Battery isolation
switch is activated.
Electrical
malfunction.
Check that battery isolation switch is deactivated (in the Power On
position). Refer to
7.3 Using the Battery Isolation Switch, page 48
Contact Provider/Invacare for Service.
Power indicator off - even after
recharging.
Battery isolation
switch is activated.
Check that battery isolation switch is deactivated (in the Power On
position). Refer to
Using the Battery Isolation Switch, page 1
.
Have terminals cleaned.