9. Connect the left motor cables
G
to the power module
P
as noted during the battery removal procedure.
Fig. 10-8 Connecting the motor cable
10. If tie-wraps were removed, tie-wrap the cables as noted
during the battery removal procedure.
11. Install the rear shroud. Refer to
the Rear Shroud Assembly , page 75
.
12. Switch the battery isolation switch to the ON position.
Refer to
7.3 Using the Battery Isolation Switch, page 48
.
13. If Installing NEW batteries—The Battery Synchronization
Procedure MUST be performed within 24 hours of
powering on the wheelchair. Otherwise the battery gauge
will not display accurate battery charge levels. The Battery
Synchronization Procedure can be found in the LiNX
service manual and must be performed by a provider or
qualified technician.
10.5.6 Replacing Battery Cables
7/16 inch (6 pt) Box Torque Wrench
84
60101916-A