to contact both battery terminals at the same time.
CAUTION!
Risk of Damage
Use of wrong battery type or size may cause damage.
-DO NOT force a battery into place.
-Batteries with molded straps or flanges that interfere
with proper battery placement should not be used.
-Use batteries listed in this manual for proper
replacement.
Failure to use the correct battery size and/or voltage
may cause damage to your wheelchair and give you
unsatisfactory performance. The warranty and
performance specifications contained in this manual
are based on the use of deep cycle gel cell batteries.
Invacare strongly recommends their use as the power
source for this unit.
Carefully read battery/battery charger information prior to
installing, servicing or operating your wheelchair.
1. Position battery on ground/flat surface as shown below.
2. Visually inspect the battery to ensure proper polarity:
FOR WHEELCHAIRS USING GP24 BATTERIES
Batteries with terminal configuration (POSITIVE
A
on the
right and NEGATIVE
B
on the left) as shown below MUST be
used. Batteries that have the reverse terminal configuration
MUST not be used - otherwise injury and damage may occur.
PROPER BATTERIES TO USE
DO NOT USE THIS TYPE OF
BATTERY
7.4.7 Instructions on Using the Batteries
CAUTION!
Risk of Damage
Draining your batteries to extremely low levels of
charge can cause battery damage. To avoid damaging
the batteries:
-Charge the batteries regularly.
-Never drain your batteries completely.
60101916-A
53
7 Controls System