72
iQue
®
Hardware Manual: Part Number 10554 Rev C., iQue Maintenance: Managing Plate Models
Calibrating a Plate Model: Gen 1 Probe Assembly
1.
Click
Device > Manage Plate Models
.
2.
Highlight the Plate Model to be calibrated and click
Calibrate
.
3.
This will bring up the Plate Model Calibration wizard. Follow the wizard instructions carefully.
4.
Place the plate being calibrated on the deck and click
Next
.
5.
Use the Vertical Movement controls to lower the probe to the proper sampling depth. The recommended
sampling depth is as close to the bottom of the well as possible without touching.
6.
Click
Finish
to save the new well depth offset.
Содержание iQue SCREENER
Страница 1: ...iQue SCREENER hardware manual October 2014...
Страница 2: ......
Страница 11: ...5 iQue Hardware Manual Part Number 10554 Rev C Introduction Declaration of Conformity...
Страница 12: ...6 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Introduction...
Страница 16: ...10 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Using This Manual...
Страница 22: ...16 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Safety...
Страница 34: ...28 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C iQue Operations Cover Page...
Страница 46: ...40 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C iQue Operations Daily QC...
Страница 50: ...44 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C iQue Maintenance Cover Page...
Страница 64: ...iQue Hardware Manual iQue Maintenance 58 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 90: ...84 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Special Topics Cover Page...
Страница 110: ......