52
iQue
®
Hardware Manual: Part Number 10554 Rev C., iQue Maintenance: Changing the Probe and Tubing
3.
Draw the loose end of the tubing up out of the
Detector area
4.
Working in the main sampling area, draw the
tubing through the sampler area pass-through.
Pull firmly for the tubing stop to come through.
Detach the Old Tubing
1.
In the Pump area, press the pump head release
trigger to unlock the pump head.
2.
In the Detector area, disconnect the tubing from
the detector inlet.