34
iQue
®
Hardware Manual: Part Number 10554 Rev C., iQue Operations: Daily Startup
NOTE
If the iQue detector takes longer than 5 minutes to reach the Ready state,
this indicates either an improper shutdown or possible fluidics issues such as
clogging.
5.
Perform Daily QC as described in the
6.
Attach the sampler tubing to the Detector Inlet.
7.
Click
Prime
and allow the system to prime for 3 minutes. Verify that the sample slugs are well formed and adjust
the peristaltic pump head tension if needed. See
.
Содержание iQue SCREENER
Страница 1: ...iQue SCREENER hardware manual October 2014...
Страница 2: ......
Страница 11: ...5 iQue Hardware Manual Part Number 10554 Rev C Introduction Declaration of Conformity...
Страница 12: ...6 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Introduction...
Страница 16: ...10 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Using This Manual...
Страница 22: ...16 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Safety...
Страница 34: ...28 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C iQue Operations Cover Page...
Страница 46: ...40 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C iQue Operations Daily QC...
Страница 50: ...44 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C iQue Maintenance Cover Page...
Страница 64: ...iQue Hardware Manual iQue Maintenance 58 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 90: ...84 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Special Topics Cover Page...
Страница 110: ......