48
iQue
®
Hardware Manual: Part Number 10554 Rev C., iQue Maintenance: Standard Cleaning Procedures
Daily Clean
The Daily Clean is performed
daily, typically at the end of the
day during the shutdown
procedure. Unlike the other
cleaning cycles, this clean is
done with the tubing attached
and it cleans the entire fluidics
path from probe to detector.
In the default configuration
(recommended) the Daily
Clean runs 5 minutes of
Decontamination Solution
followed by 5 minutes of Cleaning Solution followed by 10 minutes of water. It is possible to customize this, but be
sure to keep the overall ratio of Decon to Clean to water the same. Never run Decontamination Solution without
following it with Cleaning Solution and water.
If the
Shutdown after Clean
option is selected, the Detector shutdown procedure will be done after the main
cleaning cycle. This adds approximately 15 minutes to the clean and powers off the detector.
Quick Clean
The Quick Clean is a short procedure that cleans the detector inlet,
flow cell, and fluidics paths in the detector. It is performed as needed,
for example, when Daily QC shows a higher than normal event count or
when working with samples prone to aggregation.
To perform a Quick Clean, select the
Quick Clean
Clean Type. Place a
tube of Cleaning Solution on the detector inlet and follow the wizard
instructions. This clean takes about 10 minutes.
Long Clean
The Long Clean is a longer procedure used to clean and decontaminate
the detector inlet, flow cell, and fluidic paths. It is performed as needed,
for example, prior to shutdown at the end of the week.
To perform a Long Clean, select the
Long Clean
Clean Type. Place a tube
of Decontamination Solution on the detector inlet and follow the wizard
instructions. This clean takes about 15 minutes.
Содержание iQue SCREENER
Страница 1: ...iQue SCREENER hardware manual October 2014...
Страница 2: ......
Страница 11: ...5 iQue Hardware Manual Part Number 10554 Rev C Introduction Declaration of Conformity...
Страница 12: ...6 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Introduction...
Страница 16: ...10 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Using This Manual...
Страница 22: ...16 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Safety...
Страница 34: ...28 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C iQue Operations Cover Page...
Страница 46: ...40 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C iQue Operations Daily QC...
Страница 50: ...44 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C iQue Maintenance Cover Page...
Страница 64: ...iQue Hardware Manual iQue Maintenance 58 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 90: ...84 THIS PAGE INTENTIONALLY LEFT BLANK iQue Hardware Manual Part Number 10554 Rev C Special Topics Cover Page...
Страница 110: ......