background image

- 12 -

Figure 4

.  Time Response Selection Jumper Location

TABLE II.  Alarm2 Output Configuration (units with S2 option)

S1 Position

NC2 - COM

contacts

Relay2 State

LED1

Flow Rate

Liquid Level

N/A

Closed

de-energized

N/A

Above set point

Liquid

N/A

Open

energized

N/A

Below set point

Air

The response time of the Model 400 can be changed between two settings: “

Slow

” and “

Fast

.”  This

change is made by moving the jumper shunt at JP7 (see Figure 4).  The unit will usually come from the
factory set in the “

Slow

” position.  Notice that in the “

Slow

” position, the response is sluggish but noise-

free.  The fast response position will cause any alarming to be accelerated, however, it will allow flow
noise to be present in the instrument’s output signal.

In general, when the output signal is being monitored, the “

Slow

” position is best.  When the unit is used

strictly as a flow switch, it may be useful to use the “

Fast

” setting for gas flow or level monitoring

systems, and the “

Slow

” position for high velocity liquid flow systems.  Figure 5 shows typical response

differences between the two settings.

3.3 GENERAL INFORMATION

The Model 400 instrument is compensated for a wide range of both ambient and flowing media
temperatures.  However, abrupt changes in the temperature of the flowing material can cause the
instrument output to deviate from the true representation of flow rate, which in turn could lead to an
inappropriate tripping of the relay or a delay in reading a loss of flow.  A proper reading is obtained only
when the transducer is in thermal equilibrium with the material.  Typically, a 20°F abrupt change in
temperature may require 40 seconds to stabilize.  Temperature ramps should be kept below 1°C/minute.

In general, the heater used in the transducer does not develop enough power to cause damage to the
system in the absence of
flow.  This includes those
used in liquids even if the
line becomes empty and is
filled with air.  

Rheotherm

instruments have additional
circuitry which prevents
overheating in the event of
the loss of flow.  This
thoroughly protects the
instrument from damaging
itself.  Although it is not
required, it is recommended
that the instrument power be
shut off when the instrument
is not in use.  Any special
cases concerning unique
requirements may be
discussed in 

SECTION 6,

Содержание Rheotherm 400

Страница 1: ...its satisfaction to have been thus defective this warranty being expressly in lieu of all other warranties express or implied and all other obligation or liabilities on Intek s part The purchaser will...

Страница 2: ...UT CURVE 11 SECTION 4 MAINTENANCE 12 4 1 GENERAL MAINTENANCE 12 4 2 FIELD CALIBRATION 12 4 3 SPARE PARTS 13 4 4 TROUBLE SHOOTING 13 TABLE III TROUBLE SHOOTING GUIDE 13 SECTION 5 CUSTOMER SERVICE 14 5...

Страница 3: ...utput or equipped with two adjustable SPDT relay switches and no flow monitor output Key features of the Model 400 Rheotherm flow switches flow monitors are No moving parts There are no rotating trans...

Страница 4: ...ature signal into a temperature compensated and scaled flow output signal The signal for most cases is proportional to the logarithm of flow rate Intek also utilizes a proprietary electronic time cons...

Страница 5: ...N 2 INSTALLATION Large fluid temperature transients can result in false alarms Consult factory for recommendations to minimize the impact of thermal transients These instructions cover installation ca...

Страница 6: ...ing Ifcleaningisneeded use non residue solventandwipe dry If the sensor has a connector box keep moisture out Make sure the lid is tightly sealed and if supplied the gasket is in place Seal conduit li...

Страница 7: ...All dual probe transducers NPT 2I BF 2I are installed so that the two probes are side by side across the fluid stream Never rotate the electronics housing If this occurs the transducer can be damaged...

Страница 8: ...ment of the instrument calibration cal potentiometer after installation See SECTION 4 2 For liquid measurement systems using high pressure gas to force flow the effects of the absorbedgasmustbeconside...

Страница 9: ...d except for 115 to 230 Vac conversions To select either 24 Vac or high voltage 115 or 230 Vac ensure that power has been disconnected Then remove the electronics module by removing the modular sensor...

Страница 10: ...pulling wire pot the conduit near the enclosure if there is any possibility of them trapping water or moisture The lid of the enclosure should be on tight enough to make a good seal against the gaske...

Страница 11: ...o about forty seconds depending on where the level adjust is set Apply power only after reading and adhering to instructions in SECTION 2 of this manual 3 2 1 Adjustment of the ALARM1 flow switch setp...

Страница 12: ...10...

Страница 13: ...Trip2 potentiometer counter clockwise until the relay de energizes This is the alarm condition C If RELAY2 is de energized adjust Trip2 slowly clockwise just until the relay energizes D If the relay c...

Страница 14: ...nces between the two settings 3 3 GENERAL INFORMATION The Model 400 instrument is compensated for a wide range of both ambient and flowing media temperatures However abrupt changes in the temperature...

Страница 15: ...nit can be found after SECTION 6 For an uncalibrated unit with a 4 20mA option a typical curve is shown in Figure 6 Check SECTION 6 for the corresponding full scale value for this instrument Flow can...

Страница 16: ...ts which it operates The joints of all intrusive probes tips should be inspected for wear and corrosion 4 2 FIELD CALIBRATION If the unit was purchased with a 4 20mA output and calibrated at the facto...

Страница 17: ...ON PROBABLE CAUSE REMEDY Flow trip level continually drifting downward with constant flow Coating forming on wetted surface of transducer 1 Clean transducer periodically 2 Adjust span potentiometer cl...

Страница 18: ...and have questions please call our experienced engineers for assistance In many cases we can solve a problem over the phone Please provide as complete a description as possible of the problems encoun...

Страница 19: ...50 60 Hz G 200 240 Vac 50 Hz G other Maximum Pressure G X Full Scale Flow G Analog Output G Clamped current supplying 4 20mA flow output G none Relay Output s G ALARM1 SPDT 10A res 125 Vac G ALARM2 S...

Страница 20: ...16...

Отзывы: