background image

- 15 -

6.

In general, the curve shapes do not change if the zero and span values are set properly.  If the
instrument is being re-ranged, simply use a new full scale value and normalize the output
curve to 100%.  If extreme nonlinearities exist, such as that of a laminar to turbulent flow
regime transition, a new curve will need to be constructed.  Furthermore, if a transition region
exists and the instrument must operate over a range of temperatures, several curves must be
constructed at fixed temperature steps throughout the temperature range to assure best
accuracy.

4.3 SPARE PARTS

There are no normally  recommended spare  parts to stock.  The  transducer and electronics boards are
a matched set and therefore are not interchangeable with others.  Should a spare be needed, a complete
unit should be ordered and stocked.  Spare fuses are ½A fast acting Wickmann fuse, part number
3730500041 or equivalent for 115/230 Vac usage and 1A slow blow Wickmann fuse part number
3741100041 for 24V usage.

4.4 TROUBLE SHOOTING

TABLE III.  TROUBLE SHOOTING GUIDE

OBSERVATION

PROBABLE CAUSE

REM EDY

Flow trip level continually drifting
downward with constant flow.

Coating forming on wetted surface of
transducer.

1.

Clean transducer periodically.

2.

Adjust span potentiometer
clockwise until layer build-up
stabilizes.

After flow switch has been operating
properly:
Relay incorrectly  trips with flow
above or below trip level and cannot
be adjusted using instructions in

SECTION 3.2

.

1.

Dirty electrical contacts.

2.

Bad electronic component

1.

Inspect electronics module and
its socket for corrosion.  Clean
and replace.

2.

Contact factory.

Relay cannot be made to trip by
adjusting trip adjustment
potentiometer(s).

1.

Flow rate out of range of
instrument.

2.

Flow media change.

1.

Refer to SECTION 4.2 on
calibration to re-range the
instrument.

2.

Contact factory.

Relay trips on and off, or flow signal
is very noisy but flow is constant and
steady.

1.

Flow noise is being amplified
by fast response circuitry.

2.

Instrument has been readjusted
or instrument is defective.

1.

M ove time response jumper to
“SLOW .”

2.

Contact factory.

LED is not lit and relay stays in
alarm condition (N.C. contact is
made).

1.

No power to electronics.

2.

Blown fuse.

1.

Check incoming voltage.

2.

Replace fuse with appropriate
ampere and time delay rating.

3.

Return to factory. 

Содержание Rheotherm 400

Страница 1: ...its satisfaction to have been thus defective this warranty being expressly in lieu of all other warranties express or implied and all other obligation or liabilities on Intek s part The purchaser will...

Страница 2: ...UT CURVE 11 SECTION 4 MAINTENANCE 12 4 1 GENERAL MAINTENANCE 12 4 2 FIELD CALIBRATION 12 4 3 SPARE PARTS 13 4 4 TROUBLE SHOOTING 13 TABLE III TROUBLE SHOOTING GUIDE 13 SECTION 5 CUSTOMER SERVICE 14 5...

Страница 3: ...utput or equipped with two adjustable SPDT relay switches and no flow monitor output Key features of the Model 400 Rheotherm flow switches flow monitors are No moving parts There are no rotating trans...

Страница 4: ...ature signal into a temperature compensated and scaled flow output signal The signal for most cases is proportional to the logarithm of flow rate Intek also utilizes a proprietary electronic time cons...

Страница 5: ...N 2 INSTALLATION Large fluid temperature transients can result in false alarms Consult factory for recommendations to minimize the impact of thermal transients These instructions cover installation ca...

Страница 6: ...ing Ifcleaningisneeded use non residue solventandwipe dry If the sensor has a connector box keep moisture out Make sure the lid is tightly sealed and if supplied the gasket is in place Seal conduit li...

Страница 7: ...All dual probe transducers NPT 2I BF 2I are installed so that the two probes are side by side across the fluid stream Never rotate the electronics housing If this occurs the transducer can be damaged...

Страница 8: ...ment of the instrument calibration cal potentiometer after installation See SECTION 4 2 For liquid measurement systems using high pressure gas to force flow the effects of the absorbedgasmustbeconside...

Страница 9: ...d except for 115 to 230 Vac conversions To select either 24 Vac or high voltage 115 or 230 Vac ensure that power has been disconnected Then remove the electronics module by removing the modular sensor...

Страница 10: ...pulling wire pot the conduit near the enclosure if there is any possibility of them trapping water or moisture The lid of the enclosure should be on tight enough to make a good seal against the gaske...

Страница 11: ...o about forty seconds depending on where the level adjust is set Apply power only after reading and adhering to instructions in SECTION 2 of this manual 3 2 1 Adjustment of the ALARM1 flow switch setp...

Страница 12: ...10...

Страница 13: ...Trip2 potentiometer counter clockwise until the relay de energizes This is the alarm condition C If RELAY2 is de energized adjust Trip2 slowly clockwise just until the relay energizes D If the relay c...

Страница 14: ...nces between the two settings 3 3 GENERAL INFORMATION The Model 400 instrument is compensated for a wide range of both ambient and flowing media temperatures However abrupt changes in the temperature...

Страница 15: ...nit can be found after SECTION 6 For an uncalibrated unit with a 4 20mA option a typical curve is shown in Figure 6 Check SECTION 6 for the corresponding full scale value for this instrument Flow can...

Страница 16: ...ts which it operates The joints of all intrusive probes tips should be inspected for wear and corrosion 4 2 FIELD CALIBRATION If the unit was purchased with a 4 20mA output and calibrated at the facto...

Страница 17: ...ON PROBABLE CAUSE REMEDY Flow trip level continually drifting downward with constant flow Coating forming on wetted surface of transducer 1 Clean transducer periodically 2 Adjust span potentiometer cl...

Страница 18: ...and have questions please call our experienced engineers for assistance In many cases we can solve a problem over the phone Please provide as complete a description as possible of the problems encoun...

Страница 19: ...50 60 Hz G 200 240 Vac 50 Hz G other Maximum Pressure G X Full Scale Flow G Analog Output G Clamped current supplying 4 20mA flow output G none Relay Output s G ALARM1 SPDT 10A res 125 Vac G ALARM2 S...

Страница 20: ...16...

Отзывы: