background image

- 6 -

The probe transducer units, particularly when they are used as flow monitors, will give the
most accurate results with a straight run of pipe upstream and down stream of the probe.
Straight run length of 20 diameters upstream and 10 diameters downstream is recommended.

2.

TU or TUL (nonintrusive) — TU or TUL transducers particularly require special care in
handling and installing to avoid damage to sensor tube stubs.

.

 CAUTION:

  TU and TUL  transducers are made with thin-walled tubing — use care when

installing.

All TU and TUL transducers should have straight line input and output sections, typically 20
pipe diameters on the inlet and 6 to 10 diameters on the outlet.  If installed vertically, the flow
should be going up through the sensor.  Connection in the line is via compression fittings, hose
with clamp, threaded fittings or flanges, whichever is appropriate.  Care must be taken not to
transmit a twisting force through the transducer's midsection.  The TU or TUL transducer,
whether flanged or not, must not be used to pull other piping together or to make up angular
mismatch of fittings.  The integral electronics box should never be rotated for any reason.

Flow stream conditioning must also be considered to maximize meter performance.  Avoid
upstream protrusions and short distance straight runs.  Flow pulsations, such as those created
by metering pumps, may cause the instrument to differ from the factory calibration.
Furthermore, if the flow is varied by stroke and by pump speed adjustment, the indication will
most likely be non-repeatable.  If you are using a pump of this type, it is recommended that
a pulsation dampening device be used to provide smooth continuous flow.  A second choice
would require readjustment of the instrument calibration (cal) potentiometer after installation
(See 

SECTION 4.2

).

For liquid measurement systems using high pressure gas to force flow, the effects of the
absorbed gas must be considered.  In these cases, sudden pressure drops upstream of the sensor
such as line size expansions, control valves, and pressure dropping regulators must be avoided.
Sudden pressure drops can cause the absorbed gas to release into the liquid, making the flow
sporadic and difficult to measure.  Control valves should be placed down stream of the sensor.

Fluid temperatures other than ambient require special attention.  Thermal gradients from one
end of the transducer to the other (for TU and TUL transducers), as well as along the radius
of the connection pipe, are undesirable.  Therefore, effective insulation should be installed
around the inlet and outlet straight line runs.  Gradients which may exist in the line further up
stream can be removed if an insulated elbow is installed in the line prior to entering the
straight line portion of the plumbing.  Metallic support braces for the sensor or adjoining
plumbing can act as a heat sink and cause operational problems in high temperature
applications.  The support braces should be thermally isolated from the line to avoid large heat
conduction effects.

If the transducer is for use above 75

/

C, it will have an extension arm between the transducer

and the electronics module enclosure.  Free air should be allowed to flow around the extension
arm and electronics enclosure to keep the electronics cool.  The extension arm can be insulated
up to one-third of its length from the transducer body.  Proper thermal control is vital to
accurate meter performance.  Non-uniform heat tracing, relay on/off temperature controllers
and oscillating proportional type control should always be avoided.  Steam traced lines with
good pressure regulation or properly tuned proportional temperature control systems are

Содержание Rheotherm 400

Страница 1: ...its satisfaction to have been thus defective this warranty being expressly in lieu of all other warranties express or implied and all other obligation or liabilities on Intek s part The purchaser will...

Страница 2: ...UT CURVE 11 SECTION 4 MAINTENANCE 12 4 1 GENERAL MAINTENANCE 12 4 2 FIELD CALIBRATION 12 4 3 SPARE PARTS 13 4 4 TROUBLE SHOOTING 13 TABLE III TROUBLE SHOOTING GUIDE 13 SECTION 5 CUSTOMER SERVICE 14 5...

Страница 3: ...utput or equipped with two adjustable SPDT relay switches and no flow monitor output Key features of the Model 400 Rheotherm flow switches flow monitors are No moving parts There are no rotating trans...

Страница 4: ...ature signal into a temperature compensated and scaled flow output signal The signal for most cases is proportional to the logarithm of flow rate Intek also utilizes a proprietary electronic time cons...

Страница 5: ...N 2 INSTALLATION Large fluid temperature transients can result in false alarms Consult factory for recommendations to minimize the impact of thermal transients These instructions cover installation ca...

Страница 6: ...ing Ifcleaningisneeded use non residue solventandwipe dry If the sensor has a connector box keep moisture out Make sure the lid is tightly sealed and if supplied the gasket is in place Seal conduit li...

Страница 7: ...All dual probe transducers NPT 2I BF 2I are installed so that the two probes are side by side across the fluid stream Never rotate the electronics housing If this occurs the transducer can be damaged...

Страница 8: ...ment of the instrument calibration cal potentiometer after installation See SECTION 4 2 For liquid measurement systems using high pressure gas to force flow the effects of the absorbedgasmustbeconside...

Страница 9: ...d except for 115 to 230 Vac conversions To select either 24 Vac or high voltage 115 or 230 Vac ensure that power has been disconnected Then remove the electronics module by removing the modular sensor...

Страница 10: ...pulling wire pot the conduit near the enclosure if there is any possibility of them trapping water or moisture The lid of the enclosure should be on tight enough to make a good seal against the gaske...

Страница 11: ...o about forty seconds depending on where the level adjust is set Apply power only after reading and adhering to instructions in SECTION 2 of this manual 3 2 1 Adjustment of the ALARM1 flow switch setp...

Страница 12: ...10...

Страница 13: ...Trip2 potentiometer counter clockwise until the relay de energizes This is the alarm condition C If RELAY2 is de energized adjust Trip2 slowly clockwise just until the relay energizes D If the relay c...

Страница 14: ...nces between the two settings 3 3 GENERAL INFORMATION The Model 400 instrument is compensated for a wide range of both ambient and flowing media temperatures However abrupt changes in the temperature...

Страница 15: ...nit can be found after SECTION 6 For an uncalibrated unit with a 4 20mA option a typical curve is shown in Figure 6 Check SECTION 6 for the corresponding full scale value for this instrument Flow can...

Страница 16: ...ts which it operates The joints of all intrusive probes tips should be inspected for wear and corrosion 4 2 FIELD CALIBRATION If the unit was purchased with a 4 20mA output and calibrated at the facto...

Страница 17: ...ON PROBABLE CAUSE REMEDY Flow trip level continually drifting downward with constant flow Coating forming on wetted surface of transducer 1 Clean transducer periodically 2 Adjust span potentiometer cl...

Страница 18: ...and have questions please call our experienced engineers for assistance In many cases we can solve a problem over the phone Please provide as complete a description as possible of the problems encoun...

Страница 19: ...50 60 Hz G 200 240 Vac 50 Hz G other Maximum Pressure G X Full Scale Flow G Analog Output G Clamped current supplying 4 20mA flow output G none Relay Output s G ALARM1 SPDT 10A res 125 Vac G ALARM2 S...

Страница 20: ...16...

Отзывы: