- 3 -
5.
Keep transducer wetted surfaces clean and free of permanent layer build-up. A layer build-up
can cause the sensor to read less than the actual flow rate and can cause false relay trips.
6.
Do not exceed pressure limits of the tube or fittings.
7.
Maintain a thermally stable environment (short-term) for the transducer and adjacent line.
(
See SECTION 2 — INSTALLATION
.) Large fluid temperature transients can result in
false alarms. Consult factory for recommendations to minimize the impact of thermal
transients.
These instructions cover installation, calibration and maintenance of
Rheotherm
meters in standard
configurations. Any special information pertaining to your unit is covered under CUSTOM
INFORMATION (SECTION 6). Time should be taken to carefully read these instructions prior to
installation of the equipment. Should any questions arise or problems occur, call Intek for immediate
assistance.