Intech Instruments IN-P Series Скачать руководство пользователя страница 8

 

8

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

_ _ _ AVERAGING PARAMETERS

 

scrolls across the display 

and   toggles with 

SKIP

. Press  

P

  to skip to 6.5, or the  

   

button and then  

P

  to 

ENTER

 averaging setup. 

_ _ _ AVE SAMPLES

 

scrolls across the display and toggles 

with the currently selected averaging. Use the  

  and  

   

buttons to alter the signal averaging and then press  

 

Increasing the number of samples will stabilise measurement, 

but it will also slow down response rates. 

_ _ _ AVE WINDOW

 

scrolls across the display and toggles with 

the currently selected averaging window value.  
Use the  

  and  

  buttons to alter the signal averaging  

window, and then press  

 

       If your input signal contains large noise spikes then you can 

increase the size of averaging window to ensure that these pulses are still averaged. However, increasing the 

averaging window too far will reduce the ability of the indicator to respond quickly to real changes in input signal. 

Setting the averaging window to zero will turn off the window mode and give continuous averaging as per the se-

lected averaging samples. 

 

 

6.5 

 

Analogue output setup: 

 

If your indicator does not have this option installed then you will not view this section - setup will continue at 6.6. 

 

_ _ _ ANALOG OUTPUT SETUP

 

scrolls across display and toggles with 

SKIP

.  

Press  

P

  to skip to 6.6, or the  

  button and then  

P

  to 

ENTER

 analogue output setup. 

_ _ _ LOW SCALE VALUE FOR ANALOG OUTPUT

 

scrolls across the display and toggles with the current  

 

selection. Use the  

  and  

  buttons to enter your cal low position, and then press  

This sets the display value for cal low (as at 6.5E). 

_ _ _ HIGH SCALE VALUE FOR ANALOG OUTPUT

 

scrolls across the display and toggles with the current  

 

selection. Use the  

  and  

  buttons to enter your cal high position, and then press  

 

This sets the display value for cal high (as at 6.5F). 

_ _ _ CALIBRATE ANALOG OUTPUT?

 

scrolls across the display and toggles with 

SKIP

.  

 

If you do not wish to calibrate your analogue output, press  

P

  now. 

 

If you would like to calibrate your analogue output: 

 

Set the analogue output board jumper in the correct position (see Section 4) and connect a mA or volt meter 

across the analogue output connector (see 5.3). 

 

Press the  

  button to select 

ENTER

 and then  

P

  to enter calibration mode.

 

6.4 

 

Averaging: 

 

Your indicator has input signal averaging, guaranteeing stable measurement. If the input exceeds the averaging window 

value it will not average, ensuring fast response. 

 

 

 

 

 

 

If you selected AUTO in 6.3A:

 

_ _ _ APPLY LOW INPUT SIGNAL   –

 

 

 

 

ENTER LOW DISPLAY VALUE

 

scrolls across the display.  

 

Apply the required low input signal to the indicator. Using the  

  and  

  

buttons, enter your low end display value. 

Then press  

P

  to accept.

 

 

_ _ _ APPLY HIGH INPUT SIGNAL   –

 

 

 

 

ENTER HIGH DISPLAY VALUE

 

scrolls across the display.  

 

Apply the required high input signal to the indicator. Using the  

  and  

  

buttons, enter your high end display  

value. Then press  

. If calibration is successful, the indicator will return to the operational display.

 

 

_ _ _ CALIBRATION FAILED

 

A calibration failure results when the indicator cannot detect any change in input signal during the  calibration 

procedure. After viewing this message you will be redirected to the operational display. Check your input  

signal and connections, and then repeat the process.

 

If you selected MANUAL in 6.3A:

 

_ _ _ ENTER DISPLAY VALUE FOR (X)

 

scrolls across. (X=0mA/4mA/0V, depending on your selection in 6.2C.) 

Using the  

  and  

  

buttons, enter your display value for the selected low input signal. Then press  

.

 

 

_ _ _ ENTER DISPLAY VALUE FOR (X)

 scrolls across. (X=20mA/2V/10V, depending on your selection in 

6.2C.) Using the  

  and  

  

buttons, enter your display value for the selected high input signal. Then press  

P

  to 

accept the new calibration values and return to the operational display.

 

 

 

 

 

 

 

Содержание IN-P Series

Страница 1: ...IN P Multifunction Process Indicator Installation Guide...

Страница 2: ...tted Page 6 5 5 Wire your serial port if fitted Page 6 5 6 Wire your function pins if required Page 7 5 7 Power up your indicator Page 7 SECTION 6 Setup Calibration Page 7 6 1 Enter calibration PIN Pa...

Страница 3: ...R6 6x 5A relay outputs For PRC1 model only ANALOGUE OUTPUT RETRANSMISSION N None A 1 x 4 20mA 0 10V Analogue Output COMMUNICATIONS Not all Communication options are available for each input model type...

Страница 4: ...y will show the current VALLEY value see Section 11 E Function 2 This button is used to access the setpoint setup menu see Section 7 and the setpoint direct access menu see Section 8 SECTION 2 Specifi...

Страница 5: ...egin Before you can begin wiring the IP07 input module must be removed from the meter case so that the jumpers can be positioned for your input type Remove the plastic backing plate from the rear of t...

Страница 6: ...you do not have any relays fitted step 5 3 is skipped 5 4 Wire your analogue output if fitted Refer to 3 2I If your indicator has an analogue output fitted wire it as shown If you do not have an analo...

Страница 7: ...e the input mode in this section then the jumper plug on the input module must also be changed to match See Section 4 for more information D _ _ _ DECIMAL POINT POSITION scrolls across the display and...

Страница 8: ...lay value for cal high as at 6 5F D _ _ _ CALIBRATE ANALOG OUTPUT scrolls across the display and toggles with SKIP If you do not wish to calibrate your analogue output press P now If you would like to...

Страница 9: ...across the display and toggles with a calibration number Using the and buttons calibrate your low analogue output as required Then press P The display value is shown in internal units mA F _ _ _ CAL H...

Страница 10: ...scrolls across the display and toggles with the current selection Use the and buttons to adjust this value if required Then press P The hysteresis value defines the separation band between setpoint a...

Страница 11: ...rn to the operational display To view the minimum valley value Press the button for a second The word VALLEY appears on the display and toggles with the minimum measured process input value Press P to...

Страница 12: ...12 www intech co nz Christchurch Ph 64 3 343 0646 Auckland Ph 09 827 1930 Email sales intech co nz IN P 071216...

Отзывы: