SECTION 7. –

 

Setpoint Setup. 

 

Enter the setpoint setup mode by pressing and holding the 

 

F2

  

button for 3 

seconds. 

 

7.1 

 

Enter setpoint PIN: 

 

A

 

_ _ _ ENTER SP PIN NUMBER 

scrolls across the display and toggles with 

0

. Use the  

  and  

  buttons to 

enter your security code (factory default = 1). Then press  

. If the correct PIN is entered then the setup is started 

at 7.2. 

If an incorrect PIN number is entered, 

_ _ _ INCORRECT PIN NUMBER - ACCESS DENIED 

scrolls across the  

display and it returns to the normal operating mode. 

 

You will be given the opportunity to change your PIN number at the end of this section (7.3). 

If you have forgotten your PIN number, see Section 9. 

 

7.2 

 

Edit setpoints: 

 

A  

_ _ _ EDIT SETPOINT 

scrolls across the display and toggles with 

SKIP

. Press  

P  

to skip to 7.4, or use the  

  and 

  buttons to select a setpoint to edit: 

SP 1

SP 2

SP 3

SP 4

SP 5

 or 

SP 6

. Then press  

 

SP 3-6 are not available for models with only 2 relays installed. SP 5-6 are not available for models with only 4 relays installed.

 

_ _ _ SP VALUE 

scrolls across the display and toggles with the last setpoint value entered.  

Using the  

  and  

  buttons, adjust the display value at which the setpoint will activate. Then press  

_ _ _ SP ACTIVATION 

scrolls across the display and toggles with the last selected setpoint activation.  

Use the  

  and  

  buttons to select 

ABOVE 

or 

BELOW

 and then press  

ABOVE - relay turns on above the setpoint value and off below it. 

BELOW - relay turns on below the setpoint value and off above it. 

_ _ _ SP TYPE 

scrolls across the display and toggles with the last selected setpoint type.  

Using the  

  and  

  buttons, select either 

ALARM 

or 

CNTRL 

(Control). Then press  

_ _ _ HYSTERESIS VALUE 

scrolls across the display and toggles with the current selection.  

Use the  

  and  

  buttons to adjust this value if required. Then press  

The hysteresis value defines the separation band between setpoint activation and deactivation.  

Hysteresis will operate as per the specified type setting (see 7.2D). 

_ _ _ MAKE DELAY 

scrolls across the display and toggles with the current selection.  

Use the  

  and  

  buttons to adjust this value if required. Then press  

.  

This value is displayed in seconds with a 0.1 second resolution, and is the time delay before the relay energises. 

_ _ _ OPEN ACCESS TO SP VALUE 

scrolls across the display and toggles with the last selected direct ac-

cess setting. Use the  

  and  

  buttons to select either 

YES 

or 

NO

, and then press  

When enabled, this option allows the setpoint value to be edited directly after pressing the F2 button, without  

needing to enter a PIN number or go through all of the other options. Each setpoint can individually have this option 

enabled or disabled. 

 

 

_ _ _ TRAIL SP1

 

scrolls across the display and toggles with the current trailing setting.  

Using the  

  and  

  buttons, select either 

OFF

 or 

ON

, and then press  

If you choose ON, the selected setpoint will become an offset value, which is effectively added to the value of SP 1

_ _ _ EDIT SETPOINT 

scrolls across the display and toggles with 

SKIP

. You are now back at 7.2A.  

To edit another setpoint, follow the instructions from 7.2A-I. If you do not wish to edit another setpoint, press  

P  

now to proceed to 7.3. 

ALARM

 

The 

setpoint value 

determines setpoint activation, and 

the 

hysteresis value 

determines setpoint deactivation.

 

CNTRL (Control)

 

The 

setpoint value 

determines setpoint deactivation, and 

the 

hysteresis value 

determines setpoint reactivation.

 

    

 

     

 

If you selected SP 1 in 7.2A:

 

Skip 7.2H and continue to 7.2I now.

 

Note: 

To SKIP or ENTER values - Push 

 

 

To SELECT a menu input - Push 

 

Содержание IN-P Series

Страница 1: ...IN P Multifunction Process Indicator Installation Guide...

Страница 2: ...tted Page 6 5 5 Wire your serial port if fitted Page 6 5 6 Wire your function pins if required Page 7 5 7 Power up your indicator Page 7 SECTION 6 Setup Calibration Page 7 6 1 Enter calibration PIN Pa...

Страница 3: ...R6 6x 5A relay outputs For PRC1 model only ANALOGUE OUTPUT RETRANSMISSION N None A 1 x 4 20mA 0 10V Analogue Output COMMUNICATIONS Not all Communication options are available for each input model type...

Страница 4: ...y will show the current VALLEY value see Section 11 E Function 2 This button is used to access the setpoint setup menu see Section 7 and the setpoint direct access menu see Section 8 SECTION 2 Specifi...

Страница 5: ...egin Before you can begin wiring the IP07 input module must be removed from the meter case so that the jumpers can be positioned for your input type Remove the plastic backing plate from the rear of t...

Страница 6: ...you do not have any relays fitted step 5 3 is skipped 5 4 Wire your analogue output if fitted Refer to 3 2I If your indicator has an analogue output fitted wire it as shown If you do not have an analo...

Страница 7: ...e the input mode in this section then the jumper plug on the input module must also be changed to match See Section 4 for more information D _ _ _ DECIMAL POINT POSITION scrolls across the display and...

Страница 8: ...lay value for cal high as at 6 5F D _ _ _ CALIBRATE ANALOG OUTPUT scrolls across the display and toggles with SKIP If you do not wish to calibrate your analogue output press P now If you would like to...

Страница 9: ...across the display and toggles with a calibration number Using the and buttons calibrate your low analogue output as required Then press P The display value is shown in internal units mA F _ _ _ CAL H...

Страница 10: ...scrolls across the display and toggles with the current selection Use the and buttons to adjust this value if required Then press P The hysteresis value defines the separation band between setpoint a...

Страница 11: ...rn to the operational display To view the minimum valley value Press the button for a second The word VALLEY appears on the display and toggles with the minimum measured process input value Press P to...

Страница 12: ...12 www intech co nz Christchurch Ph 64 3 343 0646 Auckland Ph 09 827 1930 Email sales intech co nz IN P 071216...

Отзывы: