8.6.2 Replacing the starter
Fig. 105
The starters are located on the left-hand side of the front lamp
sockets.
Procedure
1.
Open the door.
2.
Switch off the unit.
3.
Remove the defect starter out of its socket by rotating it by 90°
anti-clockwise.
4.
Fit the new starter into the socket by rotating it clockwise.
NOTICE!
Fluorescent tubes contain mercury. Dispose of
fluorescent tubes according to the local statutory
regulations.
8.7 Cleaning the algae illumination
Clean the Plexiglas panel and the tubes as follows:
Procedure
1
.
Open the door.
2
.
Switch off the unit and disconnect it from the power supply.
Fig. 106
3
.
Hold onto the front edge of the Plexiglas panel with both
hands and pull it downwards uniformly off the tubes.
4
.
Clean the Plexiglas panel with a soft cloth and a mild
household cleaner.
5
.
Clean the tubes with a soft cloth and a mild household
cleaner.
6
.
Replace the Plexiglas panel with the clips by pressing it gently
onto the tubes.
Содержание Multitron Pro
Страница 1: ...Operating manual 2013 02 19 Multitron Pro Incubation Shaker Dok ID V 01 002 Original ...
Страница 177: ...Multitron Pro Incubation Shaker Appendix 2013 02 19 Page 177 of 185 12 Appendix 12 1 Declaration of Conformity ...
Страница 178: ...Multitron Pro Incubation Shaker Appendix Page 178 of 185 2013 02 19 12 2 Circuit diagram Fig 110 ...
Страница 182: ...Multitron Pro Incubation Shaker Appendix Page 182 of 185 2013 02 19 12 5 Roxtec EzEntry 4 Mini ...
Страница 183: ...Multitron Pro Incubation Shaker Appendix 2013 02 19 Page 183 of 185 ...
Страница 188: ......