Multitron Pro - Incubation Shaker
Operation
2013-02-19
Page
121
of
185
7.8 Stopping shaker table
The stopping of the shaker table can be proceeded in two ways:
Switch off parameter RPM:
Select parameter RPM with the option key.
Press SET key.
Press ON/OFF key.
Shaker table stops.
Open door:
Open door more than 30°.
Shaker table stops.
7.9 Testing the alarm function
Check the acoustic alarm and/or the correct functioning of the
alarm relay as follows:
Procedure
1.
Open the door.
2.
Set capillary thermostat to a low value, e.g. 30 °C.
3.
Shut the door.
4. Press option key to select temperature parameter.
5. The temperature parameter lights up.
6.
Press SET key.
Setpoint lights up.
7.
Press the PLUS key to set the value of the temperature
parameter above the capillary thermostat, e.g. 40 °C.
8.
Press the ON/OFF key to switch on the temperature control
system.
D
The capillary thermostat switches the heater off, when 30 °C are
reached.
The alarm is triggered with a short delay.
acoustically
optically with the warning sign and the ALARM icon.
Содержание Multitron Pro
Страница 1: ...Operating manual 2013 02 19 Multitron Pro Incubation Shaker Dok ID V 01 002 Original ...
Страница 177: ...Multitron Pro Incubation Shaker Appendix 2013 02 19 Page 177 of 185 12 Appendix 12 1 Declaration of Conformity ...
Страница 178: ...Multitron Pro Incubation Shaker Appendix Page 178 of 185 2013 02 19 12 2 Circuit diagram Fig 110 ...
Страница 182: ...Multitron Pro Incubation Shaker Appendix Page 182 of 185 2013 02 19 12 5 Roxtec EzEntry 4 Mini ...
Страница 183: ...Multitron Pro Incubation Shaker Appendix 2013 02 19 Page 183 of 185 ...
Страница 188: ......