Multitron Pro - Incubation Shaker
Interferences
Page
162
of
185
2013-02-19
Temperature is not reaching
the desired value
Heating is not
switched on
Activate the temperature parameter Operator
Setpoint value is too
low
Increase setpoint value
Operator
Capillary thermostat
is set too low (must
be at least 10 °C
above the setpoint
value)
Set capillary thermostat to higher
temperature
Operator
Heating fan is not
working
Contact Infors representative
Qualified
technician
Door is not
completely closed
Close the door properly
Operator
Top-mounted
cooling:
cooling fins are
dusty
Contact Infors representative
Qualified
technician
Side-mounted
cooling:
cooling fins are
dusty
Clean cooling fins
Operator
Содержание Multitron Pro
Страница 1: ...Operating manual 2013 02 19 Multitron Pro Incubation Shaker Dok ID V 01 002 Original ...
Страница 177: ...Multitron Pro Incubation Shaker Appendix 2013 02 19 Page 177 of 185 12 Appendix 12 1 Declaration of Conformity ...
Страница 178: ...Multitron Pro Incubation Shaker Appendix Page 178 of 185 2013 02 19 12 2 Circuit diagram Fig 110 ...
Страница 182: ...Multitron Pro Incubation Shaker Appendix Page 182 of 185 2013 02 19 12 5 Roxtec EzEntry 4 Mini ...
Страница 183: ...Multitron Pro Incubation Shaker Appendix 2013 02 19 Page 183 of 185 ...
Страница 188: ......