Side-mounted cooling system
Procedure
Fig. 101
1.
Remove the screws holding the grating.
2.
Lift off the grating.
3.
Clean the fins with a vacuum cleaner.
4.
Reattach the grating with the two screws.
Top-mounted cooling system
Procedure
Fig. 102
1.
Visually inspect the heat exchanger of the top-mounted
cooling system for dust.
2.
If there is a lot of dust, contact the manufacturer's service
center.
Coolant
Maintenance of the coolant system must be carried out by the
manufacturer's service center.
Annual check of the coolant level.
Replacement of the coolant every 3 years.