9.2 Tables of interferences
9.3 Error messages
Error messages are triggered by a mechanical or elctronical
interference on the equipment.
To additionally draw attention of the user, an interference is
displayed with the warning sign and the ERROR symbol.
The error messages disappear, when the interference was
remedied.
General interference
Interferences
Possible causes
Trouble shooting
By
Door cannot be fully opened Door is subjected to
a load
Close the door and open again
Operator
Unit is switched off
Switch the unit on
Operator
Tray lock does not engage
A foreign body is
blocking the locking
system
Remove foreign body
Operator
Foreign body is
under the table
Detach table from its axle, tilt it
upwards and remove the foreign
body.
Operator
Tray is bent or
damaged
Replace the tray
Operator
Tray is not being released
Door is not fully
open.
Press the door downwards to its
lowest position
Operator
Status display is not
illuminated
Unit is switched off
Switch the unit on
Operator
Power plug has
been disconnected
Connect the power plug
Operator
Fuse has blown
Replace fuse
Operator
Содержание Multitron Pro
Страница 1: ...Operating manual 2013 02 19 Multitron Pro Incubation Shaker Dok ID V 01 002 Original ...
Страница 177: ...Multitron Pro Incubation Shaker Appendix 2013 02 19 Page 177 of 185 12 Appendix 12 1 Declaration of Conformity ...
Страница 178: ...Multitron Pro Incubation Shaker Appendix Page 178 of 185 2013 02 19 12 2 Circuit diagram Fig 110 ...
Страница 182: ...Multitron Pro Incubation Shaker Appendix Page 182 of 185 2013 02 19 12 5 Roxtec EzEntry 4 Mini ...
Страница 183: ...Multitron Pro Incubation Shaker Appendix 2013 02 19 Page 183 of 185 ...
Страница 188: ......