Multitron Pro - Incubation Shaker
Operation
Page
130
of
185
2013-02-19
NOTICE!
Large flasks may take 20 to 30 seconds to become
detached.
Very firmly attached flasks can be detached as follows:
Procedure
1
.
Use a syringe to squirt a few drops of water onto the edge of
the cultivation flask.
2
.
Tilt the neck of the cultivation flask gently to one side.
NOTICE!
Under certain circumstances, Fernbach flasks may
be difficult to detach from the adhesive matting.
To facilitate removal, part of the adhesive matting
can be covered with the supplied protective film.
7.15 Interrupting operation
The unit can be stopped by opening the door. All functions and
parameters remain activated.
The door can only be fully opened once the table has stopped
moving.
NOTICE!
The door must be held half open while the table is
slowing down.
The door lock is only disengaged once the table
has stopped moving, provided that the door is not
subjected to a load (e.g. by something resting on
it).
To reduce operating costs, all parameters are deactivated while
the door is open.
When the door is shut, the unit restarts at a closing angle of 45°
with the preset parameters and with the interior light switched on
for 1 minute for inspection purposes.
Содержание Multitron Pro
Страница 1: ...Operating manual 2013 02 19 Multitron Pro Incubation Shaker Dok ID V 01 002 Original ...
Страница 177: ...Multitron Pro Incubation Shaker Appendix 2013 02 19 Page 177 of 185 12 Appendix 12 1 Declaration of Conformity ...
Страница 178: ...Multitron Pro Incubation Shaker Appendix Page 178 of 185 2013 02 19 12 2 Circuit diagram Fig 110 ...
Страница 182: ...Multitron Pro Incubation Shaker Appendix Page 182 of 185 2013 02 19 12 5 Roxtec EzEntry 4 Mini ...
Страница 183: ...Multitron Pro Incubation Shaker Appendix 2013 02 19 Page 183 of 185 ...
Страница 188: ......