Multitron Pro - Incubation Shaker
Installation and initial operation
2013-02-19
Page
85
of
185
There must not be any sources of electrical interference near
the unit.
All interfaces and switches on the unit must be readily
accessible.
NOTICE!
The main switch of the incubator shaker must be
completely unobstructed and readily accessible.
Ensure that the workplace is sufficiently illuminated.
Protect the operating panel against splashed water.
Protect the outside of the unit against extreme levels of dust
and dirt.
NOTICE!
Any other type of installation must first be checked
with the manufacturer and requires the written
permission of the manufacturer
6.2.2 Minimum distances
The incubator shaker must be set up as follows for operation,
handling and repair work:
NOTICE!
Main switch and cable for power supply must be
easily and freely accessible.
To provide a good circulation of air following distances are
necessary:
Right side: Minimum 8 cm
Backside: Minimum 8 cm
Top: Minimum 10 cm
The dimensions of the different models are given in the chapter
"Technical Data".
Содержание Multitron Pro
Страница 1: ...Operating manual 2013 02 19 Multitron Pro Incubation Shaker Dok ID V 01 002 Original ...
Страница 177: ...Multitron Pro Incubation Shaker Appendix 2013 02 19 Page 177 of 185 12 Appendix 12 1 Declaration of Conformity ...
Страница 178: ...Multitron Pro Incubation Shaker Appendix Page 178 of 185 2013 02 19 12 2 Circuit diagram Fig 110 ...
Страница 182: ...Multitron Pro Incubation Shaker Appendix Page 182 of 185 2013 02 19 12 5 Roxtec EzEntry 4 Mini ...
Страница 183: ...Multitron Pro Incubation Shaker Appendix 2013 02 19 Page 183 of 185 ...
Страница 188: ......