7 | Warning and error messages
INFICON
48 / 68
Contura® S400-Operating-instructions-jina91en1-09-(2001)
Type
Notification
Possible sources of
error
Remedy
W600
Calibration factor too low
Incorrect value entered
at calibration
• Repeat the calibration
Incorrect specimen
inserted
• Repeat the calibration
ZERO measurement
error
• Repeat the calibration
W601
Calibration factor too high
Incorrect value entered
at calibration
• Repeat the calibration
Incorrect specimen
inserted
• Repeat the calibration
ZERO measurement
error
• Repeat the calibration
E660
Leak in the membrane
Zero measurement with
product in the chamber
• Perform a ZERO measurement
Chamber diaphragm
leaking
• Check membrane and replace if
necessary
Hose connection (A)
leaking
• Check the hose connection (A) and
replace it if necessary
Filter leaking
• Check filter and replace if necessary
W682
Calibration volume too small
Incorrect value entered
at calibration
• Repeat the calibration
Incorrect specimen
inserted
• Repeat the calibration
ZERO measurement
error
• Repeat the calibration
Hose connection (C)
leaking
• Check the hose connection (C) and
replace it if necessary