7 | Warning and error messages
INFICON
46 / 68
Contura® S400-Operating-instructions-jina91en1-09-(2001)
Type
Notification
Possible sources of
error
Remedy
W170
Hood open
Measurement started
with chamber open
• You are measuring with an open chamber
Proximity switch not
closed
• Contact customer service
Proximity switch not
adjusted
• Contact customer service
Proximity switch
defective
• Contact customer service
W201
24 V power supply too low
Malfunction of 24V
power supply unit
• Contact customer service
Short circuit or
overload in the 24V
supply
• Contact customer service
W202
24 V power supply too high
Malfunction of 24V
power supply unit
• Contact customer service
W315
Measurement aborted
Measurement was
aborted
• Repeat the measurement
E317
Volume too large
Placed product is too
large for volume
determination
• Change the measuring mode
• Insert fewer products
• Perform a ZERO measurement
• Volume measurement recalibration
W355
Inner volume measurement
negative
Error with inner volume
measurement
• Perform a ZERO measurement
W372
Incorrect filter pressure
sensor
Error int he pressure
sensor
• Turn the power off to the device and
check if the message appears again when
you turn on the power
• Contact customer service
E500-
Contura
Pressure sensor p1 not
connected
Pressure sensor not
connected or cable
defective
• Contact customer service
E511-
Contura
Pressure sensor p2 not
connected
Pressure sensor not
connected or cable
defective
• Contact customer service