INFICON
Operation | 6
Contura® S400-Operating-instructions-jina91en1-09-(2001)
25 / 68
6.2.6.1 Determine the Measuring mode
Using the pre-set Measuring mode "Universal" you can measure packaging of
different sizes and strengths.
Undesired measurement results in the Measuring mode "Universal" are
possible
Seriously damaged packaging (gross leaks) could be shown as "Leak-proof" in
testing, because when evacuating the measuring chamber, the filling gas is
completely sucked out of the packaging.
► In order to detect larger leaks, select another measuring mode for your product from
the table below. In this way the gross leak detection will also take into consideration
the packaging.
Measuring Mode /
Packaging Size
For example
Gross leak
detection
Testing several packages
Universal
Packages of
different sizes and
strengths
All types of gas filled
packages
Deactivated
Possible
Small & Soft
Less than 500 ml
Possible. The inaccuracy of the
measurement results increases with the
number of the packages. The
measurement result is dependent on
constant volume packaging. Total volume
< 500 ml. Only same number of packaging
per measurement cycle.
Sausage packaging,
individually wrapped salami
sausage
By outer
volume
Large & Soft
Greater than 500 ml
By evaluating
the leak rate
Possible. Only same number of packaging
per measurement cycle.
Chips bag
Small & Rigid
Recommended up
to 500 ml
Possible. Only same number of packaging
per measurement cycle. Depending on the
inner volume even with large numbers.
Coffee capsules, and soft
packaging with individual
products and sufficient
cavity (nuts)
By gas filled
inner volume