10-8
Error Messages and Warnings
jinb
80
e1
-h
Op
er
at
in
g
M
anua
l
(1
50
1)
W87
Parameters are initialized in the
I•STICK! Please check the
settings!
Missing or changed parameters in the I•STICK and new software version number
were detected.
A software update was performed and one or more new parameters were
detected. In this case, the message can be acknowledged. The parameter(s) were
automatically added.
W88
Parameters are lost in the
I•STICK! Please check the
settings!
Defective or missing parameter detected in the I•STICK. The software version
number did not change.
If the warning occurs after restart again, the I•STICK cannot be written on the
wiring backplane or is defect. I•STICK on wiring backplane is defective!
The write access was interrupted. Check the settings and acknowledge the error
message.
E89
Monitoring voltage F1,F2
Voltage monitoring for fuses F1, F2 AD 24 B < 5,6 V.
Fuse F1 interface board (SSK) defective.
E90
Monitoring voltage F1,F2
Voltage monitoring for fuses F1, F2 AD 24 B < 7.8 V.
Fuse F2 interface board (SSK) defective.
W91 P_ext smaller than p_A
Pressure in the sample drops below the value of p_A, although only the test
chamber was pumped.
Sample was leaky.
Adaptation sample was leaky.
Wrong setting for p_A.
W92
Evacuation of sample takes too
long .
Time t1 > as the set evacuation time t_A.
Valve V30 does not open.
Fore pump defective
Test sample was leaky.
Wrong time setting for t_A.
Wrong setting for the evacuation pressure p_B.
W93
Filling of the test sample with
test gas takes too long.
Time t1 > as the set filling time for the sample t_C.
Valve test gas inlet V33 does not open.
Selected time for test gas inlet t_C too short.
Selected Test gas filling pressure p_C to high.
W94 Time to "Ultra" too long
Time t2 >than the time t_F to reach the measuring mode „Ultra“.
Test chamber leaky!
Sample is leaky.
Wrong setting for time t_F (time until ready for measurement "Ultra").
W95
Discharge of test gas takes too
long.
Time t1 > as the depressurizing time t_D.
Test gas pump system defect.
Valve V32 does not open.
The selected time t_D for reaching the depressurizing pressure p_D too short.
Selection of depressurizing pressure p_D wrong. Depressurizing pressure p_D >
set end pressure after the depressurizing time t_D has expired.
No.
Displayed Message
Description and possible solutions
Содержание 550-300A
Страница 6: ...1 4 Content jinb80e1 h Operating Manual 1501 ...
Страница 10: ...1 8 Operating instructions jinb80e1 h Operating Manual 1501 ...
Страница 18: ...2 8 Important safety instructions jinb80e1 h Operating Manual 1501 ...
Страница 26: ...3 8 Description of equipment jinb80e1 h Operating Manual 1501 ...
Страница 111: ...Technical Data 9 23 jinb80e1 h Operating Manual 1501 9 7 Commander mode Fig 9 14 Flow Chart ...
Страница 112: ...9 24 Technical Data jinb80e1 h Operating Manual 1501 9 8 CE Certificate Fig 9 15 ...
Страница 122: ...10 10 Error Messages and Warnings jinb80e1 h Operating Manual 1501 ...
Страница 124: ...11 2 Ordering Information jinb80e1 h Operating Manual 1501 ...