Instructions for Use and Guarantee Conditions
16
IML
-Instrumenta Mechanik Labor System GmbH
I
Parkstraße 33
I
69168 Wiesloch I
Phone +49 (0) 6222 6797-0
I
Fax +49 (0) 6222 6797-10
I
E-mail [email protected]
I
Web: www.iml.de
ENGLISH
Instruction
Exchanging the clutch disks
-
if the slide is in rear position, move it forward by 2 to 3 mm
-
remove the storage battery for safety reasons
-
remove the clutch cap
-
set torque wrench onto the adjusting nut and fix the adjusting nut by means
-
pass the counter wrench through the opening in the torque wrench and loosen
the counter nut
-
unscrew both nuts and remove worn clutch disks
-
if there is a hexagon nut, unscrew it by means of the socket wrench
-
place new clutch disks
-
screw in the adjusting nut: tighten it slowly by means of the rotating sleeve (black)
of the torque wrench until the rotating sleeve races, simultaneously fixing the spindle
by means of a screw driver
-
place the counter nut, fix the adjusting nut by means of the handle (light) of the
torque wrench and firmly tighten the counter nut by means of the counter wrench
-
re-connect the instrument with the power supply!
Note
Exchange of the clutch system (version 2)
You find a quick guide for exchanging the clutch system (version 2) in the attachment
on page 22.
Check clutch setting if the drilling advance is not regular. A clutch that has been
set too firm may lead to damages to the instrument. To ensure a long service life
of the clutch disks, you will have to see to it that it does not unneccessarily race
at the front or rear stop of the range of the recording needle!
ATTENTION