ENGLISH
Instruction
Exchanging the drilling needle
To ensure correct measuring results, you should check the
drilling needle with respect to wear after every tenth drilling.
For exchange of the needle, open the top cover and remove
the wax paper strip by pulling it upward. Next, take off the gray
security label (cf. section 'exchanging the wax paper strip')!
Forward recording needle to the intended mark by
operating the On/Off switch (+- mark).
You may see the clamping screw through the hole
on top of the paper cover.
Push red rotational direction switch in middle (neutral) position.
Now, take off the storage battery or the power pack connection cable!
Open the clamping screw with the socket screw key - direction clockwise (about 2 turns).
Now, you can release the drilling needle forward and replace it with the new one.
During positioning of the recording needle towards the mark be aware that the
drilling needle will rotate and move forward!
ATTENTION
Gray security label
Mark
Recording needle
Use the hexagon key 2.5 mm to open allen
screw (about 2 turns)!
Socket screw key
Drilling needle
Rotational direction switch for changing the drilling direction of the needle
Hole