![iM3 LAZ010 Скачать руководство пользователя страница 41](http://html1.mh-extra.com/html/im3/laz010/laz010_instructions-for-use-and-technical-description_2044810041.webp)
PG 37
EMC DECLARATION
APPENDIX D - EMC DECLARATION
The iM3 Veterinary Dental Laser
needs special precautions regarding EMC and needs to
be installed and put into service according to the EMC information provided in
the accompanying documents; Portable and mobile RF communications equipment can
affect the iM3
L
aser, even the equipment meet the requirement CISPR 11.
Guidance and manufacturer’s declaration – electromagnetic emissions
The iM3
Laser is intended for use in the electromagnetic environment specified
below. The customer or the user of the iM3 Laser
should assure that it is used in such
an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The
iM3 Laser
diode laser uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The device is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The
iM3 Laser
is intended for use in the electromagnetic environment specified
below. The customer or the user of the
iM3
diode laser should assure that it is used in
such an environment.
IMMUNITY test IEC 60601 test
level
Compliance level
Electromagnetic
environment –guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Electrical fast
transient/burst
± 2 kV for power
supply lines
± 2 kV for power
supply lines
Mains power quality should
be that of a typical