PG 28
MAINTENANCE AND SERVICE
CAUTION: Wear latex gloves when cleaning.
CAUTION: Do not use corrosive disinfectants (such as bleach).
CAUTION: Never spray to disinfect the
iM3 Laser
. Spray
disinfection may allow liquids to penetrate into the device.
5.3.5.
Cleaning Instructions for the optical window of the handpiece body
The optical window of the handpiece body is easily contaminated with dust,
resulting in a drop of the laser output. We recommend checking and cleaning
the optical window each time before use of the laser.
Use the following method for cleaning:
a)
Use compressed air to blow off the dust from the optical window.
b)
If dust remains, use Microbrush to clean the optical window.