PG 25
MAINTENANCE AND SERVICE
5. MAINTENANCE AND SERVICE
5.1.
Battery
5.1.1.
Rechargeable Battery of the Base Unit
The iM3 Veterinar
y
Laser has a built-in lithium-ion rechargeable battery.
To ensure better battery life and safety, please follow the instructions below:
1)
Lithium-ion batteries are susceptible to temperature, do not use or store
the battery at very high or low temperature. In particular, the use at high
temperature (above 60
℃
) causes irreversible damage to the battery.
2)
Lithium batteries have no memory effect. They can be charged at any
time.
3)
If the battery leaks, stop the laser and replace the battery immediately.
NOTE: To ensure the safety of the user, the
iM3 Laser
automatically stops emission when the temperature of the
lithium-ion battery is above 50
℃
.
5.1.2.
The battery of Wireless Footswitch
The wireless footswitch uses two AAA battery. When the battery is low, the
status icon on the Operation screen appears as “Low”. To replace the battery,
follow the steps below:
1)
Use a Phillips screwdriver to remove the screws and open the battery
cover of the wireless footswitch.
2)
Detach the battery terminal to remove the old battery.
3)
Replace with new AAA batteries and attach it to the terminal. Be careful
of the connection with the correct polarity!
4)
Reinstall the battery cover and screw it tightly.