![iM3 LAZ010 Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/im3/laz010/laz010_instructions-for-use-and-technical-description_2044810018.webp)
PG 14
INSTALLATION AND OPERATION
2.4.6.
Turn on the iM3 Veterinary Dental Laser To access
the SOLASE VET App:
1)
Turn on the power switch of the base unit and the green status indicator
on the front panel will light.
2)
The laser starts to self-check, which may last for more than 30 seconds.
Please wait patiently. After successful self-check, the device will ask
you to enter a 4-digit password. If you forget the password, please
contact us or our authorized service representative to reset the
password.
NOTE: The factory default password is “0000”. To change your
password, refer to section 4.4.4.
2.4.7.
Use Aiming Beam to Check the Integrity of the Delivery System
As the aiming beam passes down the same delivery system as the working
beam, it provides a good way of checking the integrity of the delivery system.
Figure 3.11 Aiming beam spot emitted from good fiber compares with that
from bad fiber
Place the distal end of the laser handpiece perpendicular to a piece of white
paper at distance of 5-10 mm. Turn on the aiming beam and adjust it to
maximum brightness. The visible light should be transmitted from the
handpiece, and a well-formed round spot should appear. If the spot is not
round in shape or its density is reduced, the beam delivery system may be
defective. Replace the laser delivery system and try again. If problems still