SOFTWARE OPERATION
powered by a battery (battery symbol) or DC power adapter (the plug
symbol). If it is powered by a battery, battery level can be counted in real
time by the number of the frame.
2)
Parameters: The adjustable parameters include Peak Power, Pulse
Duration, Pulse Interval, Handpiece Type, Emission Duration and Laser
Mode.
Peak Power: Peak power is formally defined as the maximum optical
power a laser pulse will attain. In CW mode, peak power is equal to the
average power.
Average Power: The Average Power changes automatically with Peak
Power, Pulse Duration and Pulse Interval.
Pulse Duration: Range from 10
μs
to 0.9s. 635nm cannot be set to pulse
mode.
Pulse Interval: Range from 10
μs
to 0.9s. 635nm cannot be set to pulse
mode.
Handpiece Type: Six types of handpiece are available to choose,
including 200μm FastFiber, 300μm FastFiber, 400μm FastFiber, Therapy
Handpiece, and Biostimulation Handpiece.
Emission Duration: The software records the total emission time of the
laser if this parameter is set to “0 s”. If modified, such as set to "200 s",
the total emission time is recorded and counts down to "0 s", ending
with an automatic stop of laser emission.
Laser Mode: Including CW mode and pulse mode. The Pulse Duration
and the Pulse Interval are not adjustable under CW mode. 635nm
cannot be set to pulse mode.