 Safety instructions

Read the operating instructions in full before 

starting up and follow the safety instructions.

• 

Keep the operating instructions in a place where they can be 

accessed by everyone.

• 

Ensure that only trained staff work with the device.

• 

Follow the safety instructions, guidelines, occupational health 

and safety and accident prevention regulations.

• 

The 

IKA

®

 C 3000 isoperibol calorimeter system may be used 

only in conjunction with the decomposition vessels C 6010 

or                     C 6012 for determination of the calorific values 

of solid and liquid substances to national and international 
standards (such as ISO 1928, ASTM 240, ASTM 5865, GB 
T213 und ASTM 4809).

• 

The maximum extra energy added to the decomposition ves-

sel must not exceed 

40,000

 J (select the test mass according-

ly). The permitted operating pressure of 

230 bar

 

(23 MPa)

 

must not be exceeded. The maximum permitted operating 

temperature must not exceed 

50 °C

.

• 

When burning substances containing metals, ensure that the 

maximum energy input is not exceeded!

• 

Only fill the decomposition vessel with oxygen to a maximum 

pressure of 

40 bar (4 MPa)

. Check the pressure setting on 

the pressure reducer for your oxygen supply. Perform a leak-

age test before each combustion process (see section “Leak-

tightness testing“).

• 

Some materials tend to explode when combusted (e.g. due 

to formation of peroxides), which could cause the decompo-

sition vessel to crack. 

The IKA

®

 C 3000 isoperibol calorim-

eter may not be used for testing explosive samples

.

• 

If the burning behaviour of a material is unknown, it must be 

tested before combustion in the decomposition vessel (risk of 

explosion). If you are burning 

unknown samples

, leave the 

room or 

keep your distance

 from the calorimeter.

• 

Benzoic acid may only be combusted in its pressed form! 

Flammable dust and powder must be first pressed. Oven-

dry dust and powder such as splints, hay, straw etc. explode 

when combusted! Always wet these materials first! Highly 

flammable liquids with a low vapour pressure (e.g. tetra-

methyl dihydrogen disiloxane) must not directly touch the 

cotton thread!

• 

Furthermore, toxic combustion residue in the form of gases, 

ashes or condensation, for example, is possible in the inner 

wall of the decomposition vessel.

 WARNING

Observe the accident prevention 

requirements applicable to the 

activity and the work station.

Wear your personal protective 

equipment.

• 

When handling combustion samples, combustion residue 

and auxiliary materials, please observe the relevant safety 

regulations. The following materials, for example, could pose 

a risk:

• corrosive

• highly flammable

• explosive

• bacteriologically contaminated

• toxic

• 

Please observe the relevant regulations when handling 

oxygen

.

 DANGER

Warning: oxygen as a compressed 

gas is oxidising; intensively aids 

combustion; can react violently 

with flammable materials.
Do not use oil or grease !

• 

Tubes and screwed joints for oxygen must be kept free of 

grease.

 DANGER

Combustion gases are hazardous 

to health, therefore the venting 

hose must be connected to a 

suitable gas cleaning system or 

extraction system.

• 

At the end of the work period, close the main valve for the 

oxygen supply.

• 

Perform servicing work only when the equipment is 

depressurised.

• 

When using 

stainless steel crucibles

 thoroughly check 

their condition after each experiment. If the material 

gets thinner, the crucible may catch fire and damage the 

decomposition vessel. Crucibles must not be used for more 

than 25 combustions for safety reasons.

• 

The decomposition vessel is manufactured in accordance 

with the directive for pressure equipment 2014/68/EU. This is 

indicated by the 

CE symbol

 with the ID number of the noti-

fied body. The decomposition vessel is a category III pressure 

device. The decomposition vessel has undergone an EC pro-

totype test. The CE declaration of conformity confirms that 

this decomposition vessel corresponds to the pressure device 

described in the EC prototype test certificate. The decomposi-

tion vessel has undergone a 

pressure test

 with test pressure 

of 

33 MPa 

and a leakage test with oxygen at

 3 MPa

.

• 

Refer also to the operating instructions for the C 6010/ 6012 

decomposition vessel, see chapter “maintenance of the de-

composition vessel“.

• 

Decomposition vessels are experiment autoclaves and must 

be tested by a technical expert after each use.

• 

Individual use is understood here to include a series of experi-

ments performed under roughly the same conditions in terms 

of pressure and temperature. Experiment autoclaves must be 

operated in special chambers

• 

The decomposition vessels must undergo repeated tests (in-

ternal tests and pressure tests) performed by the 

technical 

expert

. The frequency of these tests is to be determined by 

the operator on the basis of experience, type of operation 

and the material used in the decompositionvessel.

• 

The declaration of conformity becomes invalid if mechanical 

modifications are carried out to the experiment autoclaves or 

if tightness can no longer be guaranteed as a result of major 

corrosion (e.g. pitting by halogens).

• 

In particular the threads on the pressure vessel and the union 

nut are subject to considerable stress and must therefore be 

checked regularly for wear.

• 

The condition and function of the seals must be checked and 

ensured by way of a 

leakage test

 (see the operating instruc-

tions for the C 6010/ 6012 decomposition vessel, section 

“Leaktightness testing“).

Содержание C 3000 isoperibol

Страница 1: ...IKA C 3000 isoperibol 20000011772 Operating instructions EN C 3000_092016...

Страница 2: ...Device setup Touch Screen Lift Stylus for operating the touch screen Decomposition vessel Water filter Water Oxygen Interfaces Mains power supply 2...

Страница 3: ...and operating elements 11 Explanation of the screen display 11 Schematic representation of the display components 11 Weighted sample 13 Work with the device 14 Switching on the device 14 System test...

Страница 4: ...m 31 Help 31 Maintenance 31 Hardware details 32 Info 32 Cleaning 33 Cleaning the system 33 Servicing and cleaning the filter 33 Troubleshooting 34 Fault rectification 35 Checks 36 Warranty 37 Accessor...

Страница 5: ...sive bacteriologically contaminated toxic Please observe the relevant regulations when handling oxygen DANGER Warning oxygen as a compressed gas is oxidising intensively aids combustion can react viol...

Страница 6: ...ces The accessories of the system ensure customized adapta tion to laboratory tasks The system is characterized by the following features Relieves you of routine tasks through automated measurement pr...

Страница 7: ...e Transport and storage Transport and storage conditions The system must be protected against mechanical impact vibra tions dust deposits and corrosive ambient air during transporta tion and storage I...

Страница 8: ...n the chapter Safety instructions Please read these operating instructions carefully IKA consider themselves responsible for the safety reliability and performance of the appliance only when the devic...

Страница 9: ...truction IKA C 25 Connecting the oxygen supply IKA C 29 Connection to the mains power supply O2 connection tube IN max 4 MPa Insert the O2 connection tube into the calorimeter IN port until it clicks...

Страница 10: ...r USB DEVICE USB interface only for service USB HOST Interface for connection of a printer or with a USB mouse On off switch NOTICE Switch the device off at the on off switch only when directed to do...

Страница 11: ...r view based presentations of display components are described in the relevant section The contents of the screen display are often distributed over sev eral tabs You can switch between tabs at any ti...

Страница 12: ...oint in time are gath ered together as clickable buttons along the bottom edge of the screen To initiate an action it is sufficient to click on one of the discrete areas showing the relevant images or...

Страница 13: ...h as Designation Here again the entire area is available to be clicked A complete vir tual keyboard is opened Note The maximum input is 23 characters Selection box Weighted sample After the virtual ke...

Страница 14: ...pump and heater approx 30 to 45 seconds Testing the internal water circuit with the condenser and water supply plus the heater once again approx 30 to 45 seconds Testing whether the temperature of the...

Страница 15: ...e formed for example ash Often however the combustion products that form the basis of the standards are not the only products that are formed In such cases an analysis of the fuel sample and the produ...

Страница 16: ...ble absorption solution is placed in the decomposition vessel before the trial according to the rel evant standards With this absorption fluid and the combustion water the combustion gases form acids...

Страница 17: ...ecomposition vessel must then not be swapped with parts from other vessels In addition the thermal capacity depends to a small extent on the measurement procedure used The thermal capacity determined...

Страница 18: ...thread to the centre of the ignition wire using a loop 2 Weigh out the substance directly into the crucible with an accuracy of 0 1 mg If necessary put some distilled water or a solution into the dec...

Страница 19: ...e maximum permissible engery input You can also select specific types of combustion aids from the list In such cases you must enter the weight of this combustion aid so that the external energy can be...

Страница 20: ...ition vessel in the calorimeter and press Start The cover will close automatically If you accidentally press the Start button before you have hung a decomposition vessel in the calorimeter you will re...

Страница 21: ...ed calorific value and the calculated thermal capacity also appear here You can cancel the measurement at any time by clicking on Can cel After a measurement the screen switches automatically to the M...

Страница 22: ...these parameters and the relevant calculation formulae for the complete results can be found in the respective standard The screens are adapted to suit the standard that was selected When you have com...

Страница 23: ...f measurements currently displayed is prin ted not in combination with serial printer C 1 50 Delete selected measurement This deletes the selected measurement from the displayed list and also from the...

Страница 24: ...aluation button for the measurement to evaluate it for the first time or to reevalu ate it chapter Evaluation Alternatively you can use the actions provided by the menu as described above Once you hav...

Страница 25: ...calculation of the C mean value or to exclude it if it had previously been selected Then switch to the Statistics tab Statistics tab Average value standard deviation and maximum deviation relate to t...

Страница 26: ...dditional check calibrations You can use the check and warning lines to decide whether the C value has shifted and therefore a reselection and recalculation is necessary To exit this menu you must use...

Страница 27: ...ill be used for any subsequent re evaluation Otherwise the cur rent setting for the evaluation procedure will be used Measurements evaluated previously are not affected by this change Display unit Thi...

Страница 28: ...mperature at the inlet must not fall below 12 C FUT screen visible If selected the FUT screen will be shown at startup Select the menu option to connect the C 3000 to the water supply line Do not exce...

Страница 29: ...ary you can change the values for Baud rate parity data bits and stop bits Click on OK to use all the changed settings for the subsequent calorimetric procedures Click on Cancel to leave all the setti...

Страница 30: ...SB printers which the use the PCL printer language to the C 3000 We recommend the HP Business Inkjet 1200 or its successor models Network printer When you have selected a network printer you must then...

Страница 31: ...ight hand field with the arrow symbol Here you can obtain detailed information on the topic Click on the left hand field with the arrow symbol to revert to the overview of topics and if necessary sele...

Страница 32: ...urement channels various counters and statuses of accessories You cannot make any changes here The information displayed can be forwarded to Service Engineers who have access to various settings Info...

Страница 33: ...also be wiped away with an absorbent non fibrous cloth Repairs Please only send devices in for repair that have been clea ned and are free of materials which might present health hazards For this use...

Страница 34: ...ctly the Confirm button is clicked After you have taken the actions de scribed in the Help or in the operating instructions click on the Restart button to start the system test On successful completio...

Страница 35: ...e weighted sample of the sample may be insufficient The calorific value of the sample is too small use a combustion aid The oxygen pressure is insufficient The stirrer is not working The sample has no...

Страница 36: ...k using the calorimeter can be performed only once the sys tem test has been completed suc cessfully Acknowledge the error message in order to switch to the System Check status Press the Restart butto...

Страница 37: ...arge 25 pieces C 4 Quartz dish C 6 Quartz dish large C 6000 10 Spare parts kit 1000 C 6000 12 Spare parts kit 1000 For further accessories see www ika com Technical data Measuring range max J 40000 Me...

Страница 38: ...8 yes Evaluation to ASTM D5865 yes Evaluation to ISO 1928 yes Evaluation to GB T213 yes Width mm 500 Depth mm 450 Height mm 450 Weight kg 29 Permissible ambient temperature min C 20 Permissible ambien...

Страница 39: ......

Страница 40: ...IKA Werke GmbH Co KG Janke Kunkel Str 10 D 79219 Staufen Tel 49 7633 831 0 Fax 49 7633 831 98 sales ika de www ika com 20012285a...

Отзывы: