36
Display
Description
What should be done?
What happens next?
Converter alarm
High-precision temperature measu-
rement is one of the core functiona-
lities of the calorimeter.
Any error message regarding the
temperature recording brings the
current measurement to an end and
demands a system test. Further work
using the calorimeter can be perfor-
med only once the system test has
been completed successfully.
Acknowledge the error message in order to switch to the
“System Check“ status.
Press the “Restart“ button.
The temperature converter is reset and the system test
starts.
On successful completion the status switches to “Wait“
If the system test was unsuccessful, switch the C 3000
off and on again.
If the system test is unsuccessful again, please contact
IKA
®
Customer Service.
Controller alarm
The reliable regulation of the outer
vessel is a major precondition for
good quality measured values. Any
error message from the controller
brings the current measurement to
an end and demands a system test.
Further work using the calorimeter
can be performed only once the sys-
tem test has been completed suc-
cessfully.
Acknowledge the error message in order to switch to the
“System Check“ status.
Press the “Restart“ button.
The temperature converter and the controller are reset
and the system test starts.
On successful completion the status switches to “Wait“
If the system test was unsuccessful, switch the C 3000
off and on again.
If the system test is unsuccessful again, please contact
IKA
®
Customer Service.
Various major parts of the calorimeter are monitored. Information
about when the servicing intervals defined by the manufacturer
for these parts fall due is shown in the information field, with the
invitation to perform servicing actions. You must then confirm per-
formance of the servicing actions.
Click on “Confirm“ to acknowledge performance of the servicing
actions.. The acknowledgement is saved and can be checked by
the Service Engineers.
If you click on the “Help“ button you will be given instructions on
exactly what to do.
This monitoring is the precondition for a high degree of safety and
we ask you to follow the stated instructions precisely.
At the latest after 500 measurements, check the O-ring, filter etc..
Checks