25
This list shows all the decomposition vessels assigned to the ca-
lorimeter. The mark on the left in each line shows whether the
decomposition vessel is actually being used. When preparing the
sample you can only select the decomposition vessels which are
enabled. These are marked with the green tick mark
.
For reasons
of safety decomposition vessels cannot be deleted.
Each decomposition vessel is uniquely identified by the serial num-
ber. In addition the C values for the current function, the date of
last calibration and the number of ignitions are displayed.
Click on the “Check calibration for selected decomposition ves-
sels and functions“ button. Three additional tabs appear listing
the calibration, a check chart of the calibration and the calibration
statistics for the selected function and decomposition vessel.
You can select the function of the
IKA
®
C 3000 isoperibol from
a list.
”Calibration“ tab
You can select each calibration individually and click on the ˝Se-
lect“ button, or double click on the calibration to reference this
calibration in the calculation of the C mean value (or to exclude it if
it had previously been selected). Then switch to the ˝Statistics“ tab.
”Statistics“ tab
Average value, standard deviation and maximum deviation relate to
the calibration selected. Click on the button “Apply“ to load the ave-
rage value of the calibrations as the new C value.
You can also press the “Set“ button to input the new C value manu-
ally. Press the “Confirm“ button to load the value.
Press the “Print statistics“ button to print the complete statistics on
a connected printer. Press the “Delete statistics“ button to delete the
selection from the statistics that are referred to during calibration.
Click on the “Save“ button, to use all the changes for the selected
decomposition vessel and the selected function. If you click on the
“Revert“ button, you can completely undo these changes.
If you press “Done“ you save the inputs, exit this menu, and revert to
the overview of the measurements. If you press “Save“ you remain in
this menu and can select other actions (such as “Check chart“).
If you wish to change the function for subsequent measurements,
click on the ˝Change“ button to start the system check for the new
function.
Alternatively you can select a decomposition vessel from the upda-
ted list of decomposition vessels.