38
Cooling medium
water quality tap water
Type of cooling
continuous flow
Flow rate min.
l/h
60
Flow rate max.
l/h
70
Recommended flow rate at 18 °C
l/h
60
Oxygen operating pressure max
bar
40
Interface to scales
RS 232
Interface to printer
USB, RS 232 und network
Interface to PC
RS 232
Interface to ext. keyboard
yes
Oxygen filling
yes
Decomposition vessel C 6010
yes
Decomposition vessel C 6012
yes
Evaluation to ASTM D4809
yes
Evaluation to ASTM D5468
yes
Evaluation to ASTM D5865
yes
Evaluation to ISO 1928
yes
Evaluation to GB T213
yes
Width
mm
500
Depth
mm
450
Height
mm
450
Weight
kg
29
Permissible ambient temperature min.
°C
20
Permissible ambient temperature max
°C
30
Perm. humidity rel
%
80
Protection to DIN EN 60529
IP 20
Subject to technical changes!
RS 232 interface
yes
USB interface
yes
Rated voltage
V
220 - 240
Rated voltage
V
100 - 120
Frequence
Hz
50/60
Power consumption
W
1700
Fuses (220-240 V)
A
2x 6,25
Fuses (100-120 V)
A
2x 12,5
Perm. on time
continuous operation
Protection class
I
Overvoltage category
II
Contamination level
2
Operation at a terrestrial altitude
max. 2000 above sea level