![IFR 2026Q Скачать руководство пользователя страница 182](http://html1.mh-extra.com/html/ifr/2026q/2026q_operating-manual_3876146182.webp)
ACCEPTANCE TESTING
7-36
Table 7-4 RF output at +24 dBm
Carrier frequency
(MHz)
RF level min.
(dBm)
Result (dBm)
src A src B
RF level max.
(dBm)
0.03
+23
____ ____
+25
0.33
+23
____ ____
+25
60
+23
____ ____
+25
180
+23
____ ____
+25
300
+23
____ ____
+25
420
+23
____ ____
+25
540
+23
____ ____
+25
660
+23
____ ____
+25
780
+23
____ ____
+25
900
+23
____ ____
+25
1020
+23
____ ____
+25
1140
+23
____ ____
+25
1200
+23
____ ____
+25
+20 dBm
1201
+18
____ ____
+22
1260
+18
____ ____
+22
1380
+18
____ ____
+22
1500
+18
____ ____
+22
1620
+18
____ ____
+22
1740
+18
____ ____
+22
1860
+18
____ ____
+22
1980
+18
____ ____
+22
2220
+18
____ ____
+22
2340
+18
____ ____
+22
2400
+18
____ ____
+22