REMOTE OPERATION
5-26
Miscellaneous commands
(source-specific)
RPP
Reset reverse power protection
trip (short form)
:RESET
Reset RPP trip for current source.
Data type :
None
Allowed suffices :
None
Default suffix :
None
Example:
RPP:RESET
RPP:TRIPPED?
Prepares message containing information on whether the RPP Circuitry of the
current source is currently tripped in the following format:
<nr1>
(0 = not tripped, 1 = tripped)
Example:
1
RPP:COUNT?
Prepares message containing information on the number of times the RPP
Circuitry of the current source has tripped in the following format: <nr1>
Example:
3
ATTEN
[not used alone]
:LOCK
Lock the Attenuators of the current source.
:UNLOCK
Unlock the Attenuators of the current source.
Data type :
None
Allowed suffices :
None
Default suffix :
None
Example:
ATTEN:LOCK
ATTEN?
Prepares message containing information on whether the Attenuators of the
current source are locked or unlocked in the following format:
:ATTEN:<status>
where <status> is a program mnemonic indicating whether the attenuators
are locked or unlocked
Example:
:ATTEN:LOCK