REMOTE OPERATION
5-20
AM? or AM1? or AM2?
Prepares message containing information on Amplitude Modulation setting in
one of the following formats:
:AM:DEPTH <nr2>;<src>;<status>;INC <nr2>
:AM1:DEPTH <nr2>;<src>;<status>;INC <nr2>
:AM2:DEPTH <nr2>;<src>;<status>;INC <nr2>
where
<src> is a program mnemonic representing the source of the
modulation signal and <status> is a program mnemonic
indicating whether the amplitude modulation is locally ON
or OFF
Example:
:AM1:DEPTH 56.6;INT;ON;INC 5.0
AM:MODF? or AM1:MODF?
or AM2:MODF?
Prepares message containing information on AM modulation oscillator setting
in one of the following formats:
:AM:MODF:VALUE <nr2>;<shape>;INC <nr2>
:AM1:MODF:VALUE <nr2>;<shape>;INC <nr2>
:AM2:MODF:VALUE <nr2>;<shape>;INC <nr2>
where:
<shape> is a program mnemonic representing the waveform
shape
Example:
:AM:MODF:VALUE 5000.00;TRI;INC 1000.00