Tester Calibration
A calibration screen will always appear 7 days after the last field calibration.
The purpose of calibration is to (1) synchronize the units, (2) qualify (test) the
patchcords and (3) gather loss data regarding the patchcords and mated
connections.
Note: Calibration should also be done whenever there is a change of the reference
patchcords.
To calibrate the tester:
1.
Connect the channel adapters to the Display Handset and Remote Handset.
2.
Power the units on.
3.
Connect one of the two appropriate patchcords to the adapters of the DH and
RH units.
4.
From the DH Ready screen, select
Field Calibration
.
5.
From the Field Calibration screen, select
Start
on the DH.
6.
At completion of the first patchcord process, disconnect the first patchcord
from the adapters of the DH and RH and insert the second patchcord into both
of the units’ adapters.
7.
Select
Start
on the DH.
8.
At completion of the second patchcord process, disconnect the second
patchcord from one of the units and re-insert the first patchcord to the unit’s
adapter not connected to a patchcord.
The field calibration is complete and you are ready to perform tests.
If the calibration was unsuccessful, proceed to
Chapter 3, Tester Field
Calibration
. If you are unable to fix the problem, contact your local distributor or
an IDEAL INDUSTRIES service facility for assistance.
Содержание LANTEK
Страница 1: ...IDEAL INDUSTRIES LANTEK CABLE TESTERS USER S GUIDE 01 FEBUARY 2002 MANUAL PART 6510 00 5000 REV A ...
Страница 80: ...Chapter 3 Calibration Configuration and Special Operations 3 39 store up to 10 custom cables ...
Страница 116: ...Chapter 4 Autotesting 4 24 ...
Страница 124: ...Chapter 5 Analyze Testing 5 8 ...
Страница 172: ...Chapter 6 Using LANTEK REPORTER 6 48 ...
Страница 208: ...Appendix A Safety Precautions A 4 ...