Chapter 4
Autotesting
4-11
Table 4-3: Individual Autotest Reports (Continued)
Symbol
Individual Autotest Result
Fail
One or more values fail by a margin of failure greater than the unit’s
specified accuracy.
*
Fail*
One or more values fail, but the margin of failure is smaller than the unit’s
specified accuracy for the measurement.
The
fail*
result fails the cable and must be cleared up, just like a standard
fail. However,
fail*
means it has failed by a small margin and minor
adjustments in the installation may correct the error.
The effect of
pass*/fail*
reporting of individual test results is to provide a built-in
margin of safety above and below the advertised accuracy for a measurement.
Measurements which fall outside this guard band are a clear
pass
or a clear
fail
,
and the probability of the tester being in error is extremely small.
Note: If you do not want to see marginal test results, deactivate the setting in the
Preferences menu, AUTOTEST sub-menu details.