Figure 91: Completed screen with 3 options: Extraction, Relocate and Finish.
•
To continue warming, press the Extraction icon.
•
For an additional
Freeze
step - press the RELOCATE icon and select a treatment cycle.
•
To end procedure, press
Finish
.
The use of RELOCATE option is according to clinical judgment (for re-positioning the
cryoprobe in case additional freezing step is required).
Note
: You can only use the relocate option twice, thus, the system enables to Freeze only
3 locations per a given procedure.
Warning
After Extraction step, before removing the cryoprobe from the tissue, make
sure that the freeze effect has been deactivated so that the cryoprobe can
easily be removed from the tissue.
Do not force removal of the cryoprobe from the tissue as it might increase
the risk of tissue damage. Continue the Extraction step or wait for passive
thaw until the cryoprobe can be withdrawn easily
.
Warning
Do not press the
Extraction
button before the freezing protocol is
completed, unless you want to shorten the procedure due to clinical
judgment. Monitor the cryoprobe positioning during Extraction and make
sure it is still in the target tissue.
Passive Thaw
screen appears if the cryoprobe tip is too cold to disconnect the cryoprobe
from the tissue. Press
OK
after the cryoprobe is disconnected from the tissue.